470

Method

Command name

25 characters, [ MEAS Upol # ]

Name of the command.

Measurement

The two measuring modes Measurement with drift control or Meas-
urement without drift control
are selectable.

Note

A constant measured value is often only achieved after a certain time, as mixing
and possibly the reaction itself require a certain time. The response time of an
electrode can also increase with time, i.e. reaching a constant measured value
takes longer and longer. In such a case

drift-controlled measurement is par-

ticularly advisable, as the measured values are only accepted when equilibrium
has almost been achieved.

Measurement with drift control

[ on ], off

The measurement is carried out drift-controlled if this option is activated.

The measurement is stopped as soon as the

Signal drift

or the

Stop meas-

ured value

is reached or a

waiting time

has elapsed. The corresponding pa-

rameters are not visible if this option is deactivated.

Signal drift

0.01 ... [ 10.00 ] ... 99.90 µA/min

(Titrando)

0.05 ... [ 10.0 ] ... 99.9 µA/min

(Titrino)

The measured value is only accepted if the signal drift defined here has
been fallen short of.

Min. waiting time

[ 0 ] ... 999'999 s

(Titrando only)

The measured value is only accepted if the minimum waiting time has

elapsed even if the signal drift has already been reached. The drift contin-
ues to be checked as the minimum waiting time is passing.

Max. waiting time

0 ... [ 52 ] ... 999'999 s

(Titrando)

0 ... [ 52 ] ... 9'999 s

(Titrino)

If the signal drift has not yet been reached then the measured value will be

accepted when the maximum waiting time has elapsed. If no new entry for
the waiting time has been made then a suitable value for the drift will be
calculated automatically according to the following equation:

Measuring interval

0.1 ... [ 2.0 ] ... 999'999.0 s

(Titrando, increment = 0.1 s)

0.08 ... [ 2.0 ] ... 16'200 s

(Titrino, increment = 0.08 s)

Time interval for entering a measuring point in the list of measuring points.

Stop measured value

-200.0 ... 200.0 µA, [ off ]

(Titrando only)

Stops when the preset measured value has been reached since the start of
the measurement.

off

means no stop.

Measurement without drift control

[ on ], off

The measurement is carried out without drift control if this option is acti-
vated. The measurement continues until one of the stop criterions

Measur-

ing time

or

Stop measured value

is met. The corresponding parameters are

not visible if this option is deactivated.