546

Method

Device

Device name

Device name, [ not defined ]

Selection of a 774 Oven Sample Processor from the device table. Addition-

ally,

not defined

can be selected. In this case, the device name must be as-

signed by the user at the start of the method.

Device type

774 Oven Sample Processor

Display of the device type. If a device is selected under

Device name

, this

parameter is not editable any more. The device type belonging to the se-

lected device is displayed instead.
If the option

not defined

is selected as

Device name

, all device types being

able to execute the command can be selected independently from the de-
vices in the device table.

Heater

Target temperature

50 ... 250 °C, [ init.temp. ], off

Temperature at which the oven has to be heated. With

init.temp.

the oven

heater will be switched on simultaneously with the 774 Oven Sample Proc-
essor and the oven is heated up to this temperature. The initial temperature
must be defined at the 774 Oven Sample Processor and is only displayed
in the configuration of the 774 Oven Sample Processor.

Note
It is advisable to define an initial temperature if you work always at the same
temperature.

Heating time

1 ... 999 min, [ off ]

The target temperature has to be reached within this time. The heating rate

can be modified with the heating time in order to program temperature
ramps or entire temperature profiles (with several

HEATER

commands).

off

causes heating at the maximum possible heating rate (see Instructions for
Use for 774 Oven Sample Processor).

Timeout

1 ... 999 min, [ off ]

Should the target temperature not be reached within the defined

Heating

time

this waiting time starts. This parameter is only visible if a

Heating time

has been defined.

Action if timeout expires

on, [ off ]

(displayed only if

Timeout

off

)

Selection of the action to be carried out if the target temperature is not
reached after the expiry of the timeout.

Display message

[ on ], off

A message is displayed and you can choose whether you want to continue
the run or cancel it.

Stop determination

on, [ off ]

The determination is stopped and the next determination of the series is
started.

Stop determination and series

on, [ off ]

The determination as well as the series are stopped. An Error track will be
started if there is one defined.