Request - sample data request – Metrohm tiamo 1.1 Manual User Manual

Page 605

Advertising
background image

Commands

577

REQUEST - sample data request

Message

Displaying the message defined in the command

REQUEST

.

Sample position

[ 1 ] ... 999

Position of the sample on the sample rack. This number can be used for
moving to a sample position with the command

MOVE

with the

Target posi-

tion

=

Sample position

.

ID1 ... ID8

-1.0E-99 ... 1.0E+99

(for

Type

=

Number

)

100 characters

(for

Type

=

Text

)

'YYYY-MM-DD'

or

'YYYY-MM-DD hh:mm:ss'

(for

Type

=

Date/Time

)

Sample identifications.

Note

In case the sample identifications are invalid at the start of the method because
they are entered with the

REQUEST

command during the run, the option

Check at

start

in the

START

command must be switched off for these method variables.

Sample size

[ 1 ], 10 numbers

Sample size.

Sample size unit

[ g ], mg, µg, mL,µL, pieces

Unit of the sample size (text only).

Note

Sample data can be read in from a

balance or a barcode reader. The data

import must therefore be switched on in the window Run (Single determination
or Determination series) and the corresponding devices must be defined. The
dialog window

Sample data request

will be closed automatically after the data

have been received from these devices.

In case the option

Comments on modification of sample data (live)

is activated in

the Security settings, the dialog window

Modification comment sample data

(see

Subwindow Run - Single determination - Modification comment sample data) is
displayed after entering the sample data. A

Reason

and a

Comment

must be en-

tered for the modification.