Commands

409

Note

You can use the

Delay time

to calculate the maximum

Stop drift

to be used in the

following way: the size of the last added increment depends on the volume of
the attached exchange unit. With a 20 mL exchange unit (10'000 pulses per cyl-
inder volume; for KFT command is independent of the dosing drive) the smallest

possible increment is 2 µL. With a Delay time of 5 s the last 2 µL of reagent must
therefore suffice for 5 s or longer. This gives a drift of ≤ 2 µL/5 s ≤ 24 µL/min.
(The drift may be smaller than 24 µL/min as we do not know whether the last in-
crement would also have been sufficient for 10 s.) This means that if you use a
20 mL exchange unit (10'000 pulses per cylinder volume; independent of the

dosing drive) and 5 s switch-off delay then you should set a value of ≤ 24
µL/min as the stop drift.

Stop time

1 ... 999999 s

(only 720, 736, 751, 758, 784, 795, 799)

The titration is stopped if the defined stop time has elapsed. This parameter
is only shown for

Stop criterion

=

time

and

Delay time

=

inf.

.

Titration parameters

Parameters defining the run of the titration.

Command name

25 characters, [ KFT Ipol # ]

Name of the command.

Titration direction

Selection of the titration direction. If two endpoints are set the direction is al-
ready given and the following selection is ignored.

+

positive measured value alteration

[ - ]

negative measured value alteration

auto

The titration direction is determined automatically from the start measured
value and the set endpoint.

Extraction time

[ 0 ] ... 999999 s

The titration is not stopped until the extraction time has elapsed (even when

the EP has already been reached). For example, the entry of an extraction
time is a good idea for the titration of sparingly soluble samples.

Temperature

-20.0 ... [ 25.0 ] ... 150.0 °C

(Titrando)

-170.0 ... [ 25.0 ] ... 500.0 °C

(Titrino)

Manually entered titration temperature. If a temperature sensor is con-
nected and the

Temperature measurement

in the tab

General/Hardware

under

Sensor

is set to

automatic

or

continuous

then the temperature will be meas-

ured continuously.