Metrohm tiamo 1.1 Manual User Manual

Page 551

Advertising
background image

Commands

523

Device

Device name

Device name, [ not defined ]

Selection of a device from the device table. Only those devices being able

to execute the command are selectable. Additionally,

not defined

can be se-

lected. In this case, the device name must be assigned by the user at the
start of the method.

Device type

Device types, [ Titrando ]

Display or selection of the device type. If a device is selected under

Device

name

, this parameter is not editable any more. The device type belonging to

the selected device is displayed instead.
If the option

not defined

is selected as

Device name

, all device types being

able to execute the command can be selected independently from the de-
vices in the device table.

Dosing device

Dosing device

[ 1 ] ... 4

(835, 836, 842)

[ 1 ] ... 3

(855)

[ internal D0 ]

(Titrino)

external D1, external D2

(736, 751, 799)

Selection of the dosing device to be used for the dosing. All the dosing de-
vice connections (MSB) are displayed which are possible with the selected
device type.

Solution

24 characters, solution name, [ not defined ]

Input or selection of the solution defined in the solution table. Additionally,

not defined

can be selected. If intelligent exchange or dosing units are used

then a check will be made in the method sequence that the correct solution
is present in the connected dosing device and whether the type of dosing
device is correct. With non-intelligent exchange or dosing units the cylinder
volume is checked. At the start of the command the validity of the titer and

the expiry date of the selected solution will be checked as well as the GLP
test interval for the buret. If

not defined

is selected no tests will be carried

out.

Tandem dosing

on, [ off ]

(only Titrando)

If this option is enabled, the dosing is carried out without interruption with a
combination of two dosing devices so that the second dosing device is
dosing while the first one is being filled and vice versa (see STAT pH - Gen-
eral/Hardware
).

Dosing device

1 ... [ 2 ] ... 4

(835, 836, 842)

1 ... [ 2 ] ... 3

(855)

Selection of the second dosing device (exchange or dosing unit) to be
used for the dosing if the first dosing device is not available.

Solution

24 characters, solution name, [ not defined ]

(only Titrando)

Input or selection of the solution defined in the solution table. Additionally,

not defined

can be selected. If intelligent exchange or dosing units are used

then a check will be made in the method sequence that the correct solution
is present in the connected dosing device and whether the type of dosing
device is correct. With non-intelligent exchange or dosing units the cylinder

volume is checked. If

not defined

is selected no tests will be carried out.