Metrohm tiamo 1.1 Manual User Manual

Page 653

Advertising
background image

Configuration data

625

*

= any line condition, and

p

= set pulse

The output lines and bits are numbered from right to left:

Output 13 12 11 10 9 8 7 6 5 4 3 2 1 0

Bit

13 12 11 10 9 8 7 6 5 4 3 2 1 0

Example:

************1*

sets output line 1 to active (= set) which, e.g.,

would be a stop command for a connected Titrino.

************0*

sets

the line to inactive.

Note
We recommend that all the non-relevant output lines are masked with an asterisk

*

so that the conditions of these lines are not altered.

Pulse length

100 ... [ 200 ] ... 1000 ms

Duration of the transmitted pulse.

Examples for 765 Dosimat

Parameter

Bit-pattern

Function

Start Dos1

*******p******

Starts Dosimat 1 (Titrino via "activate")

Start Dos2

*****p********

Starts Dosimat 2 ( " )

Start Dos*

*****p*p******

Starts Dosimats 1 and 2 ( " )