Metrohm tiamo 1.1 Manual User Manual

Page 452

Advertising
background image

424

Method

Filling rate with different cylinders

When you use two dosing devices with different cylinder volumes

(Dosing cylinder 2 > Dosing cylinder 1) the filling rate of the larger cyl-
inder must be at the minimum:

Filling rate 2 >= Filling rate 1 • (V

Dosing cylinder 2

/ V

Dosing cylinder 1

)

Rule for dosing rate

The dosing rate must not exceed 75 % of the filling rate of the smaller
cylinder. This corresponds to the following values (at the maximum fill-

ing rate):

Cylinder volume

Max. dosing rate

Max. flow rate

Exchange unit

Dosing unit

Exchange unit

Dosing unit

1 mL

2.25 mL/min

---

ca. 130 mL/h

---

2 mL

---

5.00 mL/min

---

ca. 300 mL/h

5 mL

11.25 mL/min

12.50 mL/min

ca. 670 mL/h

ca. 750 mL/h

10 mL

22.50 mL/min

25.00 mL/min

ca. 1.3 L/h

ca. 1.5 L/h

20 mL

45.00 mL/min

50.00 mL/min

ca. 2.7 L/h

ca. 3.0 L/h

50 mL

112.50 mL/min

124.50 mL/min

ca. 6.7 L/h

ca. 7.5 L/h

Sensor

Measuring input

[ 1 ] ... 2

(835, 836, 842)

[ 1 ]

(855)

[ 1 ] ... 2, diff.

(Titrino)

Selection of the measuring input to which the sensor is connected.

Sensor

Sensor name, [ pH electrode ]

Selection of the pH sensor from the list of sensors defined in the sensor ta-
ble. The calibration data for the sensor will be adopted for the determina-
tion.

Temperature measurement

Type of temperature measuring (only Titrando):

continuous

A temperature sensor must be connected. The temperature is measured
continuously.

[ automatic ]

If a temperature sensor is connected then the temperature will be meas-

ured continuously. Otherwise the manually entered

Temperature

(tab

Titra-

tion parameters

) will used.

off

The temperature will not be measured. The

Temperature

(see Titration pa-

rameters) which has been entered manually will be used.

Stirrer

Stirrer

[ 1 ] ... 4, off

(only Titrando)

Selection of the stirrer.

off

means that no stirrer will be used.

Stirring rate

-15 ... [ 8 ] ... 15

(only Titrando)

Setting the stirrer speed. The direction in which stirring takes place changes

as the sign in front of the stirrer speed changes.

Switch off automatically

[ on ], off

(only Titrando)

If this option is enabled, the stirrer will be switched off automatically when
the command is finished.

Advertising