Metrohm tiamo 1.1 Manual User Manual

Page 593

Advertising
background image

Commands

565

Device

Device name

Device name, [ not defined ]

Selection of a device from the device table. Only those devices being able

to execute the command are selectable. Additionally,

not defined

can be se-

lected. In this case, the device name must be assigned by the user at the
start of the method.

Device type

Device types, [ Titrando ]

Display or selection of the device type. If a device is selected under

Device

name

, this parameter is not editable any more. The device type belonging to

the selected device is displayed instead.
If the option

not defined

is selected as

Device name

, all device types being

able to execute the command can be selected independently from the de-
vices in the device table.

Remote box

[ 1 ] ... 4

(only Titrando, 814, 815, 855)

Selection of the remote box at which the lines are to be scanned.

Scan lines

Input signal

Binary pattern of exactly 8 characters (0, 1,*), [ ******** ], signal template

Input of the binary pattern for the expected input signal or selecting a pre-
defined Signal pattern.
The following characters can be entered:

0

= Line inactive

1

= Line active

*

= Any line condition

The bits are always numbered from right to left:

Input 7 6 5 4 3 2 1 0

Bit

7 6 5 4 3 2 1 0

Example:

*******1

is waiting for an active input on line 0 (1 = set). For

example, this line is set by a Titrino after a titration has been completed and
the Titrino can accept a start signal again.

Note

Input lines that are of no interest or for which no defined condition can be pre-
dicted should also be masked with an asterisk (

*

).

Timeout

on, [ off ]

Maximum waiting time for the input signal.

Advertising