758

Configuration

Beaker radius

1.0 ... 100.0 mm, [ off ]

Radius of the selected special beaker on the rack. If the lift is moved to the

work position then this value will be compared with the parameter

Min.

beaker radius

(see Properties Tower #) that can be specifically defined for

each tower. If Beaker radius samples < Min. beaker radius then a cor-
responding error message will appear. With

off

the beaker radius will not be

checked.

Beaker sensor

[ Tower ], Robotic arm, off

When the selected special beaker is moved to with the command

MOVE

then the beaker sensor (

Tower

,

Robotic arm

) checks whether the special

beaker is present or not. With

off

no check will be made. For the option

Ro-

botic arm

a swing head with beaker sensor must be installed and a suitable

work position with beaker contact must be defined for the lift, as this must
move to the beaker recognition position.
With the parameter

Beaker test

in the command

MOVE

you can determine

whether the determination is to be terminated if a beaker is missing with or
without the display of a corresponding message and whether the series is

to be continued or also terminated.