Metrohm tiamo 1.1 Manual User Manual

Page 755

Advertising
background image

Subwindow Titrants/Solutions

727

Tubing parameters

Defines the length and diameter of the tubing connected to the dosing unit. The

port occupancy can also be altered. These parameters are important for carrying
out the dosing unit commands

PREP

and

EMPTY

correctly, as the volumes of the

tubing connections have to be taken into account.

Note

Default values have already been entered for the tubing parameters; these cor-
respond to the dimensions of the standard tubing supplied. If you do not make
any alterations to the tubing connections then you do not need to alter the tubing
parameters.

Dosing port 1
Port

[ Port 1 ], Port 2, Port 3, Port 4

Port to be used as Dosing port 1.

Length

0.0 ... [ 40.0 ] ... 999.9 cm

Length of tubing attached to Dosing port 1.

Diameter

0.0 ... [ 2.0 ] ... 9.9 mm

Diameter of tubing attached to Dosing port 1.

Dosing port 2
Port

Port 1, Port 2, [ Port 3 ], Port 4

Port to be used as Dosing port 2.

Length

[ 0.0 ] ... 999.9 cm

Length of tubing attached to Dosing port 2.

Diameter

0.0...[ 2.0 ]...9.9 mm

Diameter of tubing attached to Dosing port 2.

Fill port
Port

Port 1, [ Port 2 ], Port 3, Port 4

Port to be used as Fill port for aspirating the solution.

Length

0.0...[ 25.0 ]...999.9

Length of tubing attached to Fill port.

Diameter

0.0...[ 2.0 ]...9.9 mm

Diameter of tubing attached to Fill port.

Special port
Port

Port 1, Port 2, Port 3, [ Port 4 ]

Port to be used as Special port.

Length

[ 0.0 ] ... 999.9 cm

Length of tubing attached to Special port.

Diameter

0.0...[ 2.0 ]...9.9 mm

Diameter of tubing attached to Special port.