Metrohm tiamo 1.1 Manual User Manual

Page 373

Advertising
background image

Commands

345

Initial measured value

The initial measured value is determined before the other start conditions are

processed. The measured value acceptance depends on the three following pa-
rameters:

Signal drift

0.1 ... 999.0 mV/min, [ off ]

(only Titrando)

The measured value is only accepted if the drift undercuts the entered
value.

off

means that the measured value is accepted after the maximum

waiting time.

Min. waiting time

[ 0 ] ... 999999 s

(only Titrando)

The minimum waiting time is only important for drift-controlled measure-
ments. The measured value is accepted only after the minimum waiting

time has elapsed, even if the signal drift has already been achieved. The
drift continues to be checked as the minimum waiting time is passing.

Max. waiting time

0 ... [ 1 ] ... 999999 s

(only Titrando)

If signal drift has been switched off or has not yet been achieved then the
measured value will be accepted when the maximum waiting time has

elapsed.

Start volume

Start volume

[ 0.00000 ] ... 9999.99 mL

(Titrando)

[ 0.000 ] ... 999.99 mL

(Titrino)

Volume to be added before the start of the titration with the below entered

speed.

Dosing rate

0.01 ... 166.00 mL/min, [ maximum ]

(Titrando)

0.01 ... 150.00 mL/min, [ maximum ]

(Titrino)

Speed at which the start volume is to be added. The maximum dosing rate

depends on the cylinder volume of the used exchange unit or dosing unit.

Start measured value

Start measured value

-2000.0 ... 2000.0 mV, [ off ]

(only Titrando)

When the start measured value is reached the dosing of the start volume is
stopped and the next start condition is processed or the titration is started.
If the start measured value is achieved by the addition of a start volume

then the titration starts directly.

Dosing rate

0.01 ... [ 5.00 ] ... 166.00 mL/min, maximum

(only Titrando)

Speed at which the start volume is added until the start measured value is
reached. The maximum dosing rate depends on the cylinder volume of the

used exchange unit or dosing unit.

Start slope

Start slope

0 ... 999 mV/mL, [ off ]

(only Titrando)

When the start slope is reached the dosing of the start volume is stopped
and the titration is started. If the start slope is achieved by the addition of a

start volume then the titration starts directly.

Dosing rate

0.01 ... [ 5.00 ] ... 166.00 mL/min, maximum

(only Titrando)