Unattenuated beam (in front of the phantom) – Fluke Biomedical TNT 12000 User Manual

Page 100

Advertising
background image

TNT 12000

Users Manual

A-6

measurements in an attenuated beam not similar to a DN series point, the user may

calculate the actual first HVL and locate the proper correction value on the curve.

Unattenuated Beam (In Front of the Phantom)

The energy correction factors for the 96020C are determined for the unattenuated

diagnostic beam using the PTB defined DV series of beam qualities given in Table A-2.

Table A-2. Specifications for PTB Defined Unattenuated Beam Qualities

PTB

Denomination

kVp

Added Filtration

in mm AI

First HVL in mm

AI

First HVL in mm

Cu

DN40 40

2.5 1.05

0.031

DN50 50

2.5 1.42

0.045

DN60 60

2.5 1.82

0.059

DN70 70

2.5 2.45

0.081

DN80 80

2.5 3.10

0.112

DN90 90

2.5 3.60

0.126

DN100 100

2.5 4.30

0.165

DN120 120

2.5 5.40

0.231

DN150 150

2.5 1.05

0.031

Typical energy correction factor curves are obtained by dividing the calibration factor at

each beam quality by the calibration factor at a reference point and plotting the result

versus first HVL.
The typical correction factor curves for the unattenuated beam are shown below

normalized to L100 in Figure A-5 and DV70 in Figure A-6.

Typical Correction Factors - Unattenuated Beam

Normalized to L100

First HVL (mm A1)

Colrrection Factor

1

1.5

2

2.5

3

3.5

4

4.5

5

5.5

1.05
1.04
1.03
1.02
1.01

1

0.99
0.98

0.97
0.96

0.95

40

30

50

70

120

150

90

100

fct35.eps

Figure A-5. Energy Correction Factors for Model 96020C in Unattenuated Beam Normalized to L100

Advertising