TNT 12000

Users Manual

2-28

4. Use C, D, B, and A to highlight a character and push E to construct a name

that identifies the ion chamber definition. You can have a maximum of 15 characters

in the name.

5. Use C or D to highlight the Save button and push E. The ion chamber

definition name is put in memory and the Display shows the ION CHAMBER

SETUP screen.

6. Use C or D to highlight the >> button next to the Cal Factor Name label and

push E. The ALPHANUMERIC SCREEN shows in the Display.

7. Use C, D, B, and A to highlight a character and push E to construct a name

that identifies the cal factor. You can have a maximum of 15 characters in the name.

8. Use C or D to highlight the Save button and push E. The cal factor name is

put in memory and the Display shows the ION CHAMBER SETUP screen.

9. Use C or D to highlight the >> button next to the Cal Factor Value label and

push E. The NUMERIC screen shows in the Display.

fct32.bmp

10. Use A, B, C, and D to highlight a number button and push E to add the

number to the entry. You can have a maximum of seven digits in the range of -

9.999999 to +9.999999.

11. Use B or A to highlight the Save button and push E. The cal factor value is put

in memory and the Display shows the ION CHAMBER SETUP screen.

12. Use B or A to highlight << button next to the e label to enter an exponent for the cal

factor.

13. Use A, B, C, and D to highlight a number button and push E to add the

number to the entry. You can have a maximum of two digits in the range of -20 to

+20 for an exponent.

14. Use B or A to highlight the Save button and push E. The cal factor value is put

in memory and the Display shows the ION CHAMBER SETUP screen.

15. Use B or A to highlight the dropdown list and push E.
16. Use C or D to highlight one of the cal factor units and push E. Table 2-4

shows the available cal factor units.