Calibration and verification, Standard calibration, European calibration option – Fluke Biomedical TNT 12000 User Manual

Page 97: Ptb calibration option

Advertising
background image

Model 96020C Ion Chamber

Calibration and Verification

A

A-3

Average Intensity (mGy/s)

Average Intensity (R/min)

Collection Ef

ficiency

0.146

1

10

100

1000

10000

110%

100%

90%

80%

70%

60%

1.46

14.6

146

1460

Continuous

Half Wave

fct30.eps

Figure A-2. Collection Efficiency of the Model 96020C Ion Chamber

Air Density Correction ............................................. To perform air density corrections when using a calibration factor with

a reference temperature of 22

°

C, multiply the ion current by the

following correction factor, F:

F = (273.15 + T) / (295.15 x P)

Where T is the actual temperature in

°

C and P is the pressure

expressed as a fraction of a standard atmosphere (1013 hPa). For
chambers with a calibration factor normalized to 20

°

C, the

denominator is 293.15 x P.

Calibration and Verification

Calibration or calibration verification is performed by comparison of each 96020C ion

chamber with a reference chamber calibrated at either the National Institute of Standards

and Technology (NIST) or PTB.
Specifications for all PTB equivalent beam qualities are given in Table A-1 and Table A-

2 in the energy correction factor section.

Standard Calibration

The standard calibration is performed at the Fluke Biomedical, Radiation Management

Services equivalent of H60 (NIST defined as 60 kVp, first HVL of 6.0 mm Al,

homogeneity coefficient of 94). The calibration factor is normalized to 22

°

C and

1013 hPa of atmospheric pressure.

European Calibration Option

This calibration option replaces the standard calibration for units sold in Europe, or when

specifically requested by a customer. This calibration is performed at the Fluke

Biomedical, Radiation Management Services equivalent of DH70 (PTB defined as

70 kVp, first HVL of 6.77 mm Al). The calibration factor is normalized to 20

°

C and

1013 hPa of atmospheric pressure.

PTB Calibration Option

This calibration option replaces the standard calibration when PTB requirements must be

met, or when specifically requested by a customer. This option includes calibration at

both of the beam qualities listed below. Calibration factors are normalized to 20

°

C and

1013 hPa of atmospheric pressure.

Advertising