96035b ion chamber, Tnt 12000 dosemate with 96035b ion chamber, Values of the correction factor k – Fluke Biomedical TNT 12000 User Manual

Page 136

Advertising
background image

TNT 12000

Users Manual

H-4

Values of the correction factor k

Q

, for the radiation qualities as determined by

measurement at the PTB.

RQR – Series

Beam Quality

IEC 61267

Mean Energy, Air Kerma

(keV)

k

Q

(Q

n

/Q

70

)

RQR 2

26.2

1.019

RQR 3

29.0

1.012

RQR 4

32.0

1.005

RQR 5

34.8

1

RQR 6

37.8

0.996

RQR 7

41.0

0.992

RQR 8

44.2

0.989

RQR 9

50.8

0.984

RQR 10

61.2

0.979

RQA – Series

Beam Quality

IEC 61267

Mean Energy, Air Kerma

(keV)

k

Q

(Q

n

/Q

70

)

RQA 3

37.7

1.021

RQA 4

44.6

1.009

RQA 5

51.0

1

RQA 6

57.4

1.001

RQA 7

62.9

0.997

RQA 8

68.0

0.992

RQA 9

78.4

0.994

RQA 10

93.2

0.996

96035B Ion Chamber

Dimensions (Diameter x Height) ........................... 63.5 mm x 13 mm

TNT 12000 DoseMate with 96035B Ion Chamber

The following measurement and calibration nominal ranges apply.

Ranges

Dose ................................................................... 1

μ

Gy – 180 mGy

Dose Rate........................................................... 1

μ

Gy/s – 180 mGy/s

Nominal ranges of use for other factors

Conventional Diagnostic

Anode

Quality

Range of Tube Voltage

Reference

W

RQR2 - RQR10

40 - 150 kV

IEC 61267

W

RQA3 - RQA10

50 - 150 kV

IEC 61267

Mammography

Anode

Total Filtration

Range of Tube Voltage

PTB Reference

Mo

30

μ

m Mo

20 - 50 kV

MMV 20 - MMV 50

Mo

30

μ

m Mo + 2 mm Al

20 - 50 kV

MMV 20 - MMV 50

Nominal ranges of use for other factors

Factor Nominal

Range

Direction of beam incidence

Within a cone with a half angle of 5

°

to the preferred direction

Ambient temperature

15

°

C to 35

°

C

Ambient pressure

800 hPa to 1060 hPa

Relative humidity

20 % to 75 %, <20 g/m