TNT 12000

Users Manual

2-36

How to Adjust the Internal Temperature/Pressure Sensor

The Dosimeter Detector lets you override the internal measured values through an offset

value, or a specified temperature or pressure entry. Before you override the temperature

and pressure values, let the Detector equalize for a minimum of one hour. Use only

calibrated thermometers and barometers to set the temperature and pressure.
To override the temperature and/or pressure values:
Use C or D to highlight the Setup button in the Dosimeter MEASUREMENT screen

and push E. The SETUP screen shows in the Display.

fct254.png

Use C or D to highlight the T & P Sensor Adj button and push E. The TEMP &

PRES SENSOR ADJ screen shows in the Display.

fct253.png

You can override the temperature and pressure values with an offset value or a user set

value. To set an offset temperature value:
1. Use A or B to highlight the button to the right of the Internal Temperature label

and under the Offset label.

2. Push E to open the NUMERIC screen.