Fluke Biomedical TNT 12000 User Manual

Page 61

Advertising
background image

Operation

How to Use the Dosimeter Detector (TNT 12000 DoseMate)

2

2-33

The available dose units are

Roentgens (R),

Grays (Gy), and Coulombs (C).

3. Use C or D to highlight one of the dose units and push E.

Note

The TNT 12000 DoseMate PTB version only measures dose in Grays and
Coulombs.

4. Use C or D to highlight the Dose Rate dropdown box and push E. Table 2-6

shows the dose rate values.

Table 2-6. Dose Rate Values

Dose Setting

Available Dose Rates

Description

Gy/sec

Grays per second

Gy/min

Grays per minute

Gy/hr

Grays per hour

Grays (Gy)

Gy/pulse

Grays per pulse

Coulombs (C)

A

Amperes

Note

If No Ion Chamber is set, dose unit is set to C and the dose rate is set to
A. These units can not be changed.

5. Use C or D to highlight one of the dose rates and push E.

Note

The TNT 12000 PTB version only has dose rates of Gy/sec, Gy/min, Gy/hr,
and Gy/pulse when dose unit is set to Gray. Amperes is only available when
dose unit is set to Coulombs.

6. Use C or D to highlight the Frame Rate button and push E. The NUMERIC

screen shows in the Display.

Note

The Frame Rate button is only active when dose rate is set to

R/pulse or

Gy/Pulse.

fct32.bmp

7. Use A, B, C, and D to highlight a number button and push E to add the

number to the entry. Use a number between 1 and 120 with a step size of 0.1.

8. Use C or D to highlight the Save button and push E.