Appendices, A. model 96020c ion chamber, Introduction – Fluke Biomedical TNT 12000 User Manual

Page 95: Physical specifications, Appendix a

Advertising
background image

A-1

Appendix A

Model 96020C Ion Chamber

Introduction

The Model 96020C parallel-plate ion chamber is optimized for low-level intensity

measurements. The chamber has a flat energy response suitable for both attenuated and

unattenuated diagnostic beam qualities. Complete physical, electrical, and energy

response specifications are given in this appendix.

Physical Specifications

12.90 cm

1.88 cm

Direction

of incident

radiation

Diameter of active area = 11.30 cm

fct28.eps

Figure A-1. Model 96020C Ion Chamber Dimensions

Description ............................................................ Vented, parallel-plate air ionization chamber
Overall Dimensions .............................................. 12.90

±

0.03 cm diameter by 1.88

±

0.01 cm thick

Nominal Volume.................................................... 150cc, 11.30 cm diameter by 1.50 cm thick active volume
Wall Material .......................................................... Composite graphite-filled thermoplastic
Window Material.................................................... 0.76 mm thick, graphite-coated polycarbonate (lexan). A 3.5 by 4 cm

piece of 0.0025 cm thick aluminum is attached to the back of the
window to improve the energy response.