F. warnings and error messages, Introduction, Warnings and error messages – Fluke Biomedical TNT 12000 User Manual

Page 125: Appendix f

Advertising
background image

F-1

Appendix F

Warnings and Error Messages

Introduction

The Warnings and Error Messages generated by the Test Tools are in Tables F-1 through

F-4.

Table F-1. Warnings and Error Messages

Warning/Error Message

Description

XXXXXX

[1]

detector Calibration Overdue

The detector needs to be calibrated. Contact Fluke
Biomedical Service Center.

Connection lost with the detector XXXXXX

[1]

The detector has lost connection with the display or
computer. Ensure that the detector is properly
connected. Reconnect the components and select
the Connection option to reestablish the
connection.

Low Battery XXXXXX

[1]

This message appears along with a beep sound,
and the detector battery status LED glows yellow.
This happens when the detector battery has
approximately less than 20 % of charge remaining.
Connect the detector to the power source.

Low Battery Readout

This message appears along with a beep sound, the
display battery status LED glows yellow when
display battery has approximately less than 20 % of
the charge remaining. Connect the display to its ac
adapter or a USB port to continue operation.

Low Battery XXXXXX

[1]

shutting down in 2 min

This message appears along with a beep sound,
and the detector battery status LED glows red. This
happens when the detector battery has less than
10% of charge remaining. Connect the detector to
its ac adapter or a USB port to continue operation. If
this message is ignored, the detector powers off in
two minutes.