96020c ion chamber, Tnt 12000 dosemate with 96020c ion chamber – Fluke Biomedical TNT 12000 User Manual

Page 135

Advertising
background image

PTB Information

Technical Data of Type Examination by PTB

H

H-3

MAV – Series

Radiation Quality

PTB

Mean Energy (keV)

k

Q

(Q

n

/Q

70

)

MAV 25

17.6

1.006

MAV 28

18.3

1

MAV 30

18.7

0.993

MAV 35

19.4

0.983

MAV 40

19.9

0.973

MAV 50

20.7

0.960

RAV – Series

Radiation Quality

PTB

Mean Energy (keV)

k

Q

(Q

n

/Q

70

)

RAV 25

17.6

0.990

RAV 28

18.8

1

RAV 30

19.3

0.995

RAV 35

20.4

0.986

RAV 40

21.1

0.981

RAV 50

22.1

0.976

RRV – Series

Radiation Quality

PTB

Mean Energy (keV)

k

Q

(Q

n

/Q

70

)

RRV 25

15.6

1.003

RRV 28

16.3

1

RRV 30

16.7

0.997

RRV 35

17.6

0.988

RRV 40

18.2

0.973

RRV 50

19.2

0.962

96020C Ion Chamber

Dimensions (Diameter x Height) ........................... 129 mm x 18.8 mm

TNT 12000 DoseMate with 96020C Ion Chamber

The following measurement and calibration nominal ranges apply.

Ranges

Dose ................................................................... 0.1

μ

Gy – 18 mGy

Dose Rate........................................................... 0.1

μ

Gy/s – 18 mGy/s

Nominal ranges of use for other factors

Conventional Diagnostic

Anode

Quality

Range of Tube Voltage

Reference

W

RQR2 - RQR10

40 - 150 kV

IEC 61267

W

RQA3 - RQA10

50 - 150 kV

IEC 61267

Nominal ranges of use for other factors

Factor Nominal

Range

Direction of beam incidence

Within a cone with a half angle of 5

°

to the preferred direction

Ambient temperature

15

°

C to 35

°

C

Ambient pressure

800 hPa to 1060 hPa

Relative humidity

20 % to 75 %, <20 g/m