Operation

How to Use the Dosimeter Detector (TNT 12000 DoseMate)

2

2-25

fct241.png

Figure 2-14. Ion Chamber Setup Screen

If No Ion Chamber is highlighted in the ION CHAMBER SETUP screen, use D to

highlight New and push E. Go to the How to Add an Ion Chamber Definition or

Cal Factor section to add an ion chamber.
If an ion chamber definition is in the Detector:
3. Use C and D to highlight the desired ion chamber in the left frame of the screen.
4. Use B to move the highlight to the right frame of the screen.
If there are no Cal Factors for the Ion Chamber, then highlight New and push E.

See the How to Add an Ion Chamber Definition or Cal Factor section to make a cal

factor.
If there is a Cal factor, then:
5. Use C and D to highlight the desired calibration factor and push E.
6. Use B or A to highlight the Continue button and push E.
If Air Density Correction Factor (ADCF) is set to off or if it is on and the

Temperature/Pressure is configured for Internal Temperature/Pressure Compensation

then the Dosimeter Measurements screen shows on the Display. Go to the Dosimeter

Measurements section to continue the measurement procedure.