Fluke Biomedical TNT 12000 User Manual

Page 127

Advertising
background image

Warnings and Error Messages

Introduction

F

F-3

Table F-3, DoseMate Warnings and Error Messages (cont.)

Warning/Error Message

Description

“Resetting…Please Wait…” message is displayed

This message displays when you select Reset in
the Measurement Screen or when the DoseMate
dosimeter automatically resets.
NOTE: Do not make an exposure when the
“Resetting…Please Wait…” message is displayed. If
an exposure is initiated while this message is
displayed, the message will remain until the
exposure is ended.

“Temperature/Pressure Sensor Error” message is
displayed

This message displays when the Temperature or
the Pressure value sensed by the integrated
Temperature/Pressure sensor of the DoseMate is
invalid.
The Dose and Dose Rate Value measured during
the Temperature/Pressure Sensor Error is invalid.
Hence, the Manual Entry of Temperature and
Pressure is recommended. (Refer section “Air
Density Correction: External
Temperature/Pressure Mode
”)

The “Bias Voltage Error, please restart the system”
message is displayed

This message displays when the DoseMate Ion
Chamber Bias voltage is not within normal operating
range. When this occurs, the Ion Chamber bias
voltage is turned off. To restore the bias voltage, the
DoseMate must be turned off and then on again.
The Bias Voltage error may be caused by a faulty
ion chamber and/or cable. To verify, disconnect the
ion chamber and/or cable, turn the DoseMater off
and then on. The ion chamber bias voltage may be
checked from the Status screen.

Measured quantities showing HIGH

This message indicates an over range. The current
or charge from the Ion Chamber is above the
maximum.

Table F-4. mAs Detector Warnings and Error Messages

Warning/Error Message

Description

Measured quantities showing HIGH

This message indicates an over range. The
measured mA or mAs is above the maximum
current.