Diagnostic – attenuated beam (behind the phantom) – Fluke Biomedical TNT 12000 User Manual

Page 108

Advertising
background image

TNT 12000

Users Manual

B-6

Typical Correction Factors - Diagnostic Unattenuated Beam

Normalized to DV70

First HVL (mm A1)

Colrrection Factor

1

1.5

2

2.5

3

3.5

4

4.5

5

5.5

1.08

1.06

1.04

1.02

1

0.98

0.96

0.94

0.92

40

30

50

70

120

150

90

100

fct36.eps

Figure B-4. Energy Correction Factors for Model 96035C in Unattenuated Beam Normalized to DV70

All 96035B ion chambers must have an actual correction factor within

±

1.5 % of the

value shown in the curves at any point.
The numerical value shown beside each point is the kVp value of that point in the DV

series of beam qualities. Users making measurements in an unattenuated beam may use

these values to easily obtain the appropriate correction factor. For a more precise

correction factor value, the user may calculate the actual first HVL and locate the proper

correction value on the curve.

Diagnostic – Attenuated Beam (Behind the Phantom)

The energy correction factors for the 96035B in the attenuated diagnostic beam are

determined using the PTB defined DN series of beam qualities given below in Table B-2.

Table B-2. Specifications for PTB Defined Attenuated Beam Qualities

PTB

Denomination

kVp

Added Filtration

in mm AI

First HVL in mm

AI

First HVL in mm

Cu

DN40 40

6.5

2.2

0.07

DN50 50

12.5

3.4

0.123

DN60 60

18.5

5.0

0.207

DN70 70

23.5

6.2

0.289

DN80 80

29.5

7.8

0.403

DN90 90

32.5

9.0

0.501

DN100 100

36.5

10.1

0.609

DN120 120

42.5

12.0

0.839

DN150 150

52.5

14.1

1.245


Typical correction factor curves are obtained by dividing the calibration factor at each

beam quality by the calibration factor at a reference point and plotting the result versus

first HVL.
The typical correction factor curve for the attenuated beam is normalized to DH70 in

Advertising