B. model 96035b ion chamber, Introduction, Physical specifications – Fluke Biomedical TNT 12000 User Manual

Page 103: Appendix b

Advertising
background image

B-1

Appendix B

Model 96035B Ion Chamber

Introduction

The Model 96035B dual entrance window parallel-plate ion chamber has a flat response

suitable for both diagnostic radiography and mammography. One entrance window is

provided for diagnostic measurements and one for mammographic measurements. The

fully guarded, centrally located collector plate provides excellent collection efficiency.

This appendix provides complete physical, electrical, and energy response specifications

.

Physical Specifications

2.86 cm

0.71 cm

1.30 cm

Body Center

Line

Direction

of incident

radiation for

diagnostic

measurements

3.96 cm

Active Region

Center Line

Active Region

6.35 cm

DIAGNOSTIC

FOCUS

DIAGNOSTIC

FOCUS

fct29.eps

Figure B-1. Model 96035B Ion Chamber Dimensions

Description ............................................................ Vented volume, parallel-plate air ionization chamber
Overall Dimensions .............................................. 6.35

±

0.04 cm diameter by 1.30

±

0.01 cm thick

Nominal Volume.................................................... 15 cc, 3.96 cm diameter by 1.22 cm thick active volume
Wall Material .......................................................... Graphite-coated acrylic