Values of the correction factor k – Fluke Biomedical TNT 12000 User Manual
Page 137
PTB Information
Technical Data of Type Examination by PTB
H
H-5
Values of the correction factor k
Q
, for the radiation qualities as determined by
measurement at the PTB.
RQR – Series
Beam Quality
IEC 61267
Mean Energy, Air Kerma
(keV)
k
Q
(Q
n
/Q
70
)
RQR 2
26.2
1.035
RQR 3
29.0
1.023
RQR 4
32.0
1.010
RQR 5
34.8
1
RQR 6
37.8
0.991
RQR 7
41.0
0.983
RQR 8
44.2
0.976
RQR 9
50.8
0.966
RQR 10
61.2
0.956
RQA – Series
Beam Quality
IEC 61267
Mean Energy, Air Kerma
(keV)
k
Q
(Q
n
/Q
70
)
RQA 3
37.7
1.049
RQA 4
44.6
1.020
RQA 5
51.0
1
RQA 6
57.4
0.987
RQA 7
62.9
0.980
RQA 8
68.0
0.978
RQA 9
78.4
0.982
RQA 10
93.2
0.986
MMV – Series
Beam Quality PTB
Mean Energy (keV)
k
Q
(Q
n
/Q
70
)
MMV 20
13.5
1.012
MMV 25
14.9
1.004
MMV 28
15.4
1
MMV 30
15.7
0.999
MMV 35
16.3
0.995
MMV 40
16.7
0.993
MMV 50
17.3
0.990
MMH – Series
Beam Quality PTB
Mean Energy (keV)
k
Q
(Q
n
/Q
70
)
MMH20 17.3
0.984
MMH 25
18.6
0.982
MMH 28
19.3
0.981
MMH 30
19.8
0.980
MMH 35
21.0
0.980
MMH 40
22.1
0.978
MMH 50
24.2
0.974