Values of the correction factor k – Fluke Biomedical TNT 12000 User Manual

Page 137

Advertising
background image

PTB Information

Technical Data of Type Examination by PTB

H

H-5

Values of the correction factor k

Q

, for the radiation qualities as determined by

measurement at the PTB.

RQR – Series

Beam Quality

IEC 61267

Mean Energy, Air Kerma

(keV)

k

Q

(Q

n

/Q

70

)

RQR 2

26.2

1.035

RQR 3

29.0

1.023

RQR 4

32.0

1.010

RQR 5

34.8

1

RQR 6

37.8

0.991

RQR 7

41.0

0.983

RQR 8

44.2

0.976

RQR 9

50.8

0.966

RQR 10

61.2

0.956

RQA – Series

Beam Quality

IEC 61267

Mean Energy, Air Kerma

(keV)

k

Q

(Q

n

/Q

70

)

RQA 3

37.7

1.049

RQA 4

44.6

1.020

RQA 5

51.0

1

RQA 6

57.4

0.987

RQA 7

62.9

0.980

RQA 8

68.0

0.978

RQA 9

78.4

0.982

RQA 10

93.2

0.986

MMV – Series

Beam Quality PTB

Mean Energy (keV)

k

Q

(Q

n

/Q

70

)

MMV 20

13.5

1.012

MMV 25

14.9

1.004

MMV 28

15.4

1

MMV 30

15.7

0.999

MMV 35

16.3

0.995

MMV 40

16.7

0.993

MMV 50

17.3

0.990

MMH – Series

Beam Quality PTB

Mean Energy (keV)

k

Q

(Q

n

/Q

70

)

MMH20 17.3

0.984

MMH 25

18.6

0.982

MMH 28

19.3

0.981

MMH 30

19.8

0.980

MMH 35

21.0

0.980

MMH 40

22.1

0.978

MMH 50

24.2

0.974