Other calibration options, Calibration accuracy, Calibration verification – Fluke Biomedical TNT 12000 User Manual

Page 98: Energy correction factors, Attenuated beam (behind the phantom)

Advertising
background image

TNT 12000

Users Manual

A-4

Unattenuated beam............................................... Calibration is performed at the Fluke Biomedical, Radiation

Management Services equivalent of DV70 (PTB defined as 70 kVp and
first HVL of 2.45 mm Al).

Attenuated beam................................................... Calibration is performed at the Fluke Biomedical, Radiation

Management Services equivalent of DH70 (PTB defined as 70 kVp
and first HVL of 6.77 mm Al).

Other Calibration Options

Optional calibration points are the Fluke Biomedical, Radiation Management Services

equivalent of L100, M50, S75, DV30, DV40, DV50, DV60, DV80, DV90, DV100,

DV120, DV150, DH40, DH50, DH60, DH80, DH90, DH100, DH120, and DH150.

Calibration factors at these points may be normalized to either 20

°

C or 22

°

C, as

specified by the customer.

Calibration Accuracy

NIST traceable ........................................................ NIST traceable reference chambers are accurate to within

±

1 % at

each beam quality. Matching between the reference and unknown
chamber is accurate to

±

1 %. Thus, NIST traceable calibrations are

accurate to within

±

2 %.

PTB traceable ......................................................... PTB traceable reference chambers are accurate to within

±

1.5 % at

each beam quality. Matching between the reference and unknown
chamber is accurate to

±

1 %. Thus, PTB traceable calibrations are

accurate to within

±

2.5 %.

Calibration Verification

NIST traceable ........................................................ For NIST traceable calibrations, a chamber will be reported to be

outside of its calibration accuracy specification when a new calibration
value differs from the old calibration value by more than

±

4 %. This

includes the

±

2 % accuracy specification plus

±

1 % for the verification

reference plus

±

1 % for the verification match.

PTB traceable ......................................................... For PTB traceable calibrations, a chamber will be reported to be

outside of its calibration accuracy specification when a new calibration
value differs from the old calibration value by more than

±

5 %. This

includes the

±

2.5 % accuracy specification plus

±

1.5 % for the

verification reference plus

±

1 % for the verification match.

Energy Correction Factors

Attenuated Beam (Behind the Phantom)

The energy correction factors for the 96020C are determined for the attenuated diagnostic

beam using the PTB defined DN series of beam qualities given in Table A-1.

Table A-1. Specifications for PTB Defined Attenuated Beam Qualities

PTB

Denomination

kVp

Added Filtration

in mm AI

First HVL in mm

AI

First HVL in mm

Cu

DN40 40

6.5

2.2

0.07

DN50 50

12.5

3.4

0.123

DN60 60

18.5

5.0

0.207

DN70 70

23.5

6.2

0.289

DN80 80

29.5

7.8

0.403

DN90 90

32.5

9.0

0.501

DN100 100

36.5

10.1

0.609

DN120 120

42.5

12.0

0.839

DN150 150

52.5

14.1

1.245