Air density correction – Fluke Biomedical TNT 12000 User Manual

Page 62

Advertising
background image

TNT 12000

Users Manual

2-34

9. Use C or D to highlight the Temperature dropdown box and push E.

Note

If ADCF is off, the temperature option is disabled.

10. Use C or D to highlight

°

C

,

°

F, or K

and push E.

11. Use C or D to highlight the Pressure dropdown box and push E.
12. Use C or D to

highlight inHg, mmHg, mbar,

hPa

, feet, or meters

and push E.

Note

If ADCF is off, the pressure option is disabled.

13. Use C or D to highlight the Save button and push E to save the units

selections.

Air Density Correction

When Air Density Correction is set to on, the Dosimeter corrects dose and dose rate

measurements based on temperature and barometric pressure. Temperature and pressure

values can be set externally through the Display or internally by the Dosimeter

temperature and pressure sensors.
To turn the Air Density Correction Factor (ADCF) on or off:
1. Use A or B to highlight the Mode button in the Dosimeter MEASUREMENT screen

and push E. The MODE SETUP screen shows in the Display.

fct250.png

2. Use C or D to highlight the Air Density Correction dropdown box and push

E

.

3. Use C or D to highlight On or Off and push E.
To set Temperature, Pressure, and Reference Temperature in the MODE

SETUP screen, Air Density Correction must be set to On. If set to Off, all three

controls are disable in the MODE SETUP screen.