Fluke Biomedical TNT 12000 User Manual

Page 50

Advertising
background image

TNT 12000

Users Manual

2-22

How to Place the Ion Chamber for Horizontal X-ray Beams

To setup the ion chamber on the test stand, see Figure 2-11 and do the steps that follow.

X-Ray Beam

Cable Clip

Filter Tray

Rod Clamp

HVL Filter

Baseplate

Ion Chamber

Ion Chamber

Bracket

Rod Clamp

fct21.eps

Figure 2-11. Test Stand Configuration for Horizontal Tube

1. Put the short square rod into the baseplate and tighten the thumbscrew.
2. Put the rod clamp on the short square rod and move it to approximately the middle of

the rod. Tighten the rod clamp thumbscrew to the short rod.

3. Put the long square rod into the remaining rod clamp hole and tighten the rod clamp

thumbscrew. This makes a horizontal square rod.