Fluke Biomedical TNT 12000 User Manual

Page 63

Advertising
background image

Operation

How to Use the Dosimeter Detector (TNT 12000 DoseMate)

2

2-35

External Temperature/Pressure Compensation Mode

It is assumed that air density correction is set to On in the steps that follow.
1. Use C, D, A, and B to highlight the checkbox next to the User Set Temperature

label and push E to toggle the checkbox. When the box is checked, the button

to the right of the checkbox is enabled. When this checkbox is not checked, internal

temperature compensation is activated.

2. Use B to highlight the temperature button and push E to change the temperature

value. The NUMERIC screen shows in the Display.

3. Use C, D, A, and B to highlight a number and push E to add the number to the

entry.

4. When the temperature value is set, highlight the Save button and push E to set

the temperature compensation value.

5. Use C, D, A, and B to highlight the checkbox next to the User Set Pressure label

and push E to toggle the checkbox. When the box is checked, the button to the

right of the checkbox is enabled. When this checkbox is not checked, internal

pressure compensation is activated.

6. Use C, D, A, and B to highlight the Reference Temperature dropdown box and push

E

.

7. Use C or D to highlight 20 or 22 and push E. The ion chamber is calibrated at

this reference temperature. See the Ion Chamber Specifications for the highlighted

chamber in Appendix A, B, C, or D.

8. Use C or D to highlight the Continue button and push E. The mode is set

and the MEASUREMENT screen shows in the Display.

Internal Temperature/Pressure Compensation Mode

When the Dosimeter is set for internal temperature compensation, the temperature value

is set by an internal sensor. This temperature can be different than the temperature of the

ion chamber connected to the Detector. The Detector and the ion chamber must be the

same temperature as the environment before a measurement.
The Dosimeter internal temperature approximately 30 minutes after it is turned on is

typically 10 to 15

°

C above ambient temperature. At standard conditions of 22

°

C and

760 mm Hg

, a change of 1

°

C ~ 0.36 % change in ADC. Fluke Biomedical recommends

the internal temperature sensor only be used to measure temperature after the Detector is

turned on and the Dosimeter and ion chamber are the same temperature as the

environment. Then set this temperature into the Detector manually.

Note

With its internal temperature compensation, the internal pressure sensor
works in all conditions.

Advertising