Calibration and verification, Standard calibration, European calibration option – Fluke Biomedical TNT 12000 User Manual

Page 105

Advertising
background image

Model 96035B Ion Chamber

Calibration and Verification

B

B-3

Average Intensity (mGy/s)

Average Intensity (R/min)

Collection Ef

ficiency

0.146

1

10

100

1000

10000

110%

100%

90%

80%

70%

60%

1.46

14.6

146

1460

Continuous

Half Wave

fct30.eps

Figure B-2. Collection Efficiency of Model 96035C Ion Chamber

Ion Transit Time .................................................... Maximum of 0.67 ms in the center of the chamber at STP with a bias

voltage of 300 volts.

Air Density Correction.......................................... To perform air density corrections when using a calibration factor with

a reference temperature of 22

°

C, multiply the ion current by the

following correction factor, F:

F = (273.15 + T) / (295.15 x P)

Where T is the temperature in

°

C and P is the pressure expressed as

a fraction of a standard atmosphere (1013 hPa). For chambers with a
calibration factor normalized to 20

°

C, the denominator is 293.15 x P.

Calibration and Verification

Calibration or calibration verification is performed by comparison of each 96035B ion

chamber with a reference chamber calibrated at either the National Institute of Standards

and Technology (NIST) or PTB.
Specifications for all PTB equivalent beam qualities are given in the tables in the energy

correction factor section.

Standard Calibration

The standard calibration is performed at one diagnostic and one mammographic beam

quality as given below. Calibration factors are normalized to 22

°

C and 1013 hPa of

atmospheric pressure.
Diagnostic Unattenuated beam: Calibration is performed at the Fluke Biomedical,

Radiation
Management Services equivalent of L100 (NIST defined as 100 kVp, first HVL of

2.8 mm Al, homogeneity coefficient of 59).
Mammographic beam: Calibration is performed at the Fluke Biomedical, Radiation

Management Services equivalent of MoMo28 (NIST defined as 28 kVp, first HVL of

0.332 mm Al, homogeneity coefficient of 74.3).

European Calibration Option

This calibration option replaces the standard calibration for units sold in Europe, or when

specifically requested by a customer. This option includes one diagnostic and one

mammographic beam quality, as given below. Calibration factors are normalized to

20

°

C and 1013 hPa of atmospheric pressure.

Advertising