How to Use the X-ray Detector (TNT 12000WD)

2

2-9

Measurement Mode Setup

To set the X-ray measurement mode:
1. In the MEASUREMENT screen, use A or B to highlight the 2

nd

button from the left.

The name on this button shows the measurement mode selection. In Figure 2-6, the

button is named Radio which shows the Detector is set to the Radio Measurement

mode. The MODE SETUP screen in Figure 2-7 shows in the Display.

fct209.png

Figure 2-7. Mode Setup Screen

2. Use C or D to highlight one of the measurement mode buttons and push E.

The color of the button name changes to blue.

3. Use C or D to highlight the Continue button and push E to set the

measurement mode and return to the MEASUREMENT screen. The name on the

middle button changed to the name of the new measurement mode.

Note

Before the measurement mode is set, move the highlight to the Cancel
button and push
E to cancel the measurement mode change.

Highlight the Settings button in this screen to change the parameters for the

measurement mode. See the How to Change Profile Parameters section.

Measurement Mode Profile Parameters

Each measurement mode has a number of parameters that can be changed and set into the

Detector. These parameters are used to setup the Detector for the X-ray exposure and set

the units of measure. These parameters together make a special Detector profile. Table 2-

2 shows the settable profile parameters.

Table 2-2. Settable Profile Parameters

Parameter Description

Dose Units

Sets the Dose unit between

Roentgen (R) or

Gray (Gy).

[3]

Time Units
(Radio, mammo, dental only)

Sets time unit to milliseconds (msec) or pulses.

Dose Rate Units
(Fluoro only)

Sets the dose Rate units to second, minute, hour, or pulse.

[3]

% kV

[2]

(Radio, mammo, dental only)

Sets %kV to 0%, 75%, 80%, or 90%.

[1]