Ion chamber selection and setup, Ion chamber connection and placement – Fluke Biomedical TNT 12000 User Manual

Page 44

Advertising
background image

TNT 12000

Users Manual

2-16

highlight the IC button in the MEASUREMENT screen and push E. Go to the Ion

Chamber Selection and Setup section to set No Ion Chamber in the ION CHAMBER

SETUP screen.

Ion Chamber Selection and Setup

Choose an ion chamber for the measurement sensitivity requirements, X-ray beam

dimensions, beam-accessing port dimensions, and/or special function performance (e.g.

CT probe). After you pick an ion chamber, the subsequent step is to find a chamber

definition in the TNT 12000. If the definition is not found, make one.

Ion Chamber Connection and Placement

XW Warning

To prevent electric shock and damage to the ion chamber,
power off the Dosimeter Detector before you connect or
disconnect the ion chamber and triaxial cable. Although current
limited, the center pin and inner shield shell operates at 300 V
above the outer shell. High voltage is internally discharged
within a few seconds when the Detector is turned off.

W Caution

To prevent damage to the ion chamber, do not connect a BNC

type cable to the ion chamber input connector.

Note

Keep the Dosimeter Detector out of the radiation field. Only the ion
chamber must be exposed to the radiation.

To connect an ion chamber to the Dosimeter Detector:
1. Remove the dust cover from the ion chamber input connector.
2. Connect one end of the Triaxial ion chamber cable to the ion chamber connector as

shown in Figure 2-8.

Advertising