Fluke Biomedical TNT 12000 User Manual

Page 27

Advertising
background image

Introduction

Specifications

1

1-15

150 cc Ion Chamber

Units Effective

Range

[1]

Threshold

Resolution Step Size

R

10

μ

to 2

6

μ

0.1

μ

R/s

10

μ

to 2

6

μ

0.1

μ

R/min

0.5 m to 120

0.36 m

5

μ

R/hr

10 m to 7.2 k

21.6 m

0.1 m

R/frame

[2]

0.2

μ

to 33 m

0.1

μ

0.002

μ

Gy

0.1

μ

to 0.02

0.052

μ

0.001

μ

Gy/s

0.1

μ

to 0.02

0.052

μ

0.001

μ

Gy/min

5

μ

to 1.2

3.15

μ

0.05

μ

Gy/hr

0.1 m to 72

0.189 m

0.001 m

Gy/frame

[2]

0.002

μ

to 0.33 m

0.8 n

0.02 n

[1]

IEC 61674 effective range at 1 % resolution steps.

[2]

At 60 frames/s (1 to 120 frames/selectable).


150 cc Low Rate Ion Chamber

Units Effective

Range

[1]

Threshold

Resolution Step Size

R/s

2

μ

to 2

[3]

NA

0.1

μ

R/min

0.1 m to 120

[3]

NA

5

μ

R/hr

2 m to 7.2 k

[3]

NA

0.1

m

R/frame

[2]

0.04

μ

to 33 m

[3]

NA

0.002

μ

Gy/s

0.02

μ

to 0.02

[3]

NA

0.001

μ

Gy/min

1

μ

to 1.2

[3]

NA

0.05

μ

Gy/hr

0.02 m to 72

[3]

NA 0.001

m

Gy/frame

[2]

0.4 n to 0.33 m

[3]

NA

0.02

n

[1]

IEC 61674 effective range at 1 % resolution steps.

[2]

At 60 frames/s (1 to 120 frames/selectable).

[3]

Low Rate effective range at 5 % resolution steps.


Electrical Units

Units Effective

Range

[1]

Threshold

Resolution Step Size

C

1 p to 100 n

0.5 pC

0.01 p

A

1 p to 100 n

250 fA

0.01 p

[1]

IEC 61674 effective range at 1 % resolution steps.