Mas specifications – Fluke Biomedical TNT 12000 User Manual

Page 25

Advertising
background image

Introduction

Specifications

1

1-13

Filtration Correction Range

Radio/Fluro/Dental Modes .................................. 1-10 mm Al or equivalent
Mammo Mode ..................................................... 0–0.4 mm Al or equivalent added filtration (0–0.2 mmAI below 25 kV)

kV Correction Range

Radio/Fluro/Dental Modes .................................. 40 kV–150 kV
Mammo Mode: Mo/Mo........................................ 22 kV-35 kV

Exposure Time–Radiographic Modes
Range @ stated accuracy

Milliseconds ........................................................ 10–9999 ms
Pulses ................................................................. 1–999 pulses

Resolution

Milliseconds ........................................................ 0.1 ms
Pulses ................................................................. 1 pulse

Accuracy

Milliseconds ........................................................ 1 % or 0.5 ms
Pulses ................................................................. ±1 pulse

Reproducibility

Milliseconds ........................................................ 1 % or 0.5 ms
Pulses ................................................................. ±1 pulse

Elapsed Time–Fluoro Modes
Range ..................................................................... 10–9999 seconds
Resolution ............................................................... 0.1 second
Accuracy ................................................................. 1 % or 0.5 second
Average Pulse Rate – Pulsed Fluoro
Range ..................................................................... 1–999 pps (pulses per second)
Resolution ............................................................... 1 pps
Accuracy ................................................................. 1 pps
Average Pulse Width – Pulsed Fluoro
Range ..................................................................... 10–999 ms
Resolution ............................................................... 0.1 ms
Accuracy ................................................................. 1% or 0.5 ms
HVL (Half Value Layer)
Range

Radio/Fluoro/Dental Modes ................................ 1.2–10 mm Al (equivalent)
Mammo Mode ..................................................... 0.2–0.6 mm Al (equivalent)

Resolution

Radio/Fluoro/Dental Modes ................................ 0.1 mm Al (equivalent)
Mammo Mode ..................................................... 0.01 mm Al (equivalent)

Accuracy

Radio/Fluoro/Dental Modes ................................ ±10 % or 0.2 mm Al (equivalent)
Mammo Mode: Mo/Mo........................................ ± 5 % or 0.05 mm Al (equivalent)

mAs Specifications

Accuracy

With Invasive Shunt ............................................ ±2 % of rdg ±2 digits
With Non-Invasive Shunt .................................... ±3 % ±3 mA

Invasive mA/mAs Range......................................... 0.00 – 99.99 mA/mAs

100.0 – 999.9 mA/mAs
1000 – 1999 mA/mAs

Non-Invasive mA/mAs Ranges ............................... 0 – 999.9 mA/mAs

1000 – 3999 mA/mAs

Trigger Levels

Invasive Shunt .................................................... 3 mA
Non-Invasive Shunt ............................................ 6 mA

Input Jack................................................................ Male Coaxial BNC
mA/mAs Interface Cable ......................................... 152.4 cm

(60 in)

BNC Male – Female, 50 Ohm cable with strain relief