TNT 12000

Users Manual

1-12

Environmental

Operating Temperature........................................... 0

°

C to 35

°

C

(32

°

F to 95

°

F)

Storage Temperature.............................................. -35

°

C to 50

°

C

(-31

°

F to +122

°

F)

Operating Humidity ................................................. 20 % to 80 %

RH

(Non Condensing)

Wireless Range....................................................... 30 m

(100 ft)

X-ray Specifications

kVp
Units........................................................................ kVp Average (Average of peaks during a specified interval)

kVp Max (Highest peak during a specified interval)
PPV (Peak Practical Voltage)

Ranges
Radio/Fluoro/Dental Modes (W,Al) ......................... 40 kV–150 kV
Mammo Mode

Mo/Mo ................................................................. 22 kV–35 kV
Rh/Rh.................................................................. 25 kV– 49 kV
Mo/Rh ................................................................. 22 kV–40 kV
Mo/Al................................................................... 22 kV– 49 kV
Rh/Al ................................................................... 25 kV–49 kV
W/Rh ................................................................... 22 kV–39 kV
W/Ag ................................................................... 22 kV–39 kV

Resolution: .............................................................. 0.1 kV
Accuracy (including calibration uncertainty)
Radio/Fluoro/Dental Modes .................................... ±2 % or ±1 kV, whichever is greater
Mammo Mode ......................................................... ±2 % or ±0.7 kV, whichever is greater
Reproducibility ........................................................ ±1 % (standard deviation % of 5 readings)
Filtration Correction Range
Radio/Fluoro/Dental Modes .................................... 1–10 mm Al or equivalent
Mammo Mode ......................................................... 0–0.4 mm Al or equivalent added filtration (0–0.2 mm AI below 25 kV)
Dose/Exposure
Units........................................................................

Roentgens (R)

Grays (Gy)

Range .....................................................................

0.5 mR–999 R

5 mGy–999 Gy

Resolution ...............................................................

1 µR

0.01 µGy

Accuracy ................................................................. ±5 %
Reproducibility ........................................................ ±0.5 % (standard deviation % of five readings)
Filtration Correction Range

Radio/Fluoro/Dental Modes ................................ 10 mm Al or equivalent
Mammo Mode ..................................................... 0–0.4 mm Al or equivalent added filtration (0– 0.2 mm AI below 25 kV)

kV Correction Ranges

Radio/Fluoro/Dental Modes ................................ 40 kV–150 kV
Mammo Mode: Mo/Mo........................................ 22 kV–35 kV

Dose/Exposure Rate
Units........................................................................

Roentgens per hour, minute, second, pulse (R/hr, R/min, R/sec,
R/Pulse)

Grays per hour, minute, second, pulse (Gy/hr, Gy/min, Gy/sec,
Gy/Pulse)

Range .....................................................................

8 mR/s–10 R/s

70 µGy/s–100 mGy/s

130 µR/Pulse–160 mR/Pulse (@ 60 PPS)

12 µGy/Pulse–1.4 mGy/Pulse (@ 60 PPS)

Resolution ...............................................................

1 µR/s

0.01 µGy/s

0.02 µR/Pulse (@ 60 PPS)

0.2 nGy/Pulse (@ 60 PPS)

Accuracy ................................................................. ±5 %