How to change profile parameters – Fluke Biomedical TNT 12000 User Manual

Page 40

Advertising
background image

TNT 12000

Users Manual

2-12

Table 2-3. Default Parameter Values for Each Measurement Mode

Measurement Mode

Parameter

Radio Mammo Fluoro Dental

Dose Units

R (Roentgen)

R (Roentgen)

R (Roentgen)

R (Roentgen)

Time Units

msec (milliseconds) msec (milliseconds)

NA

msec (milliseconds)

Dose Rate Units

NA

NA

R/min

NA

% kV

90 %

90 %

NA

0 %

Fluoro Type

NA

NA

CONT (continuous)

NA

Target/Filter W/AI Mo/Mo W/AI W/AI

Delay

0 msec

0 msec

NA

0 msec

How to Change Profile Parameters

The four manual profiles, four auto profiles, and all user-defined profiles contain

parameters that can be changed. After parameters are changed and then used to set the

Detector, those changes redefine the profile in memory.
To change the parameters in a profile:
1. Highlight the Settings button in the MODE SETUP screen or the Edit Profile

button in the MEASUREMENT screen.

2. Push E.

The SETTING screen shows in the Display.

fct218.png

Note

The manual settings screen for the Radio mode is shown in the above
example. The banner text shows the mode selection (RADIO, MAMMO,
FLUORO, or DENTAL) and the profile (AUTO, MANUAL, DEFAULT or
user-defined). See Table 2-2 for the parameters that are available for each
measurement mode.

3. Use C or D to highlight the parameter you want to change and push E to open

the drop-down list.

4. Use C or D to highlight a value in the list and push E to set the parameter to