Fluke Biomedical TNT 12000 User Manual

Page 46

Advertising
background image

TNT 12000

Users Manual

2-18

X-Ray Beam

Filter Tray

Cable Clip

Square Rod

Ion Chamber

HVL Filter

Baseplate

Ion Chamber
Holder

fct26.eps

Figure 2-9. Test Stand Configuration for Over-Table Tubes

1. Put the square rod into the test-stand baseplate and tighten the baseplate thumbscrew.
2. Put the ion chamber bracket on the square rod. Move it along the rod until the top of

the bracket aligns with the scribe mark in the middle of the rod. Tighten the chamber

bracket thumbscrew.

3. Put the HVL filter tray on the rod. Move it along the rod until the top of the tray

aligns with the top scribe mark on the rod. Tighten the HVL filter thumbscrew.

4. Attach the ion chamber to the bracket.