Fluke Biomedical TNT 12000 User Manual
Page 47
Advertising
Operation
How to Use the Dosimeter Detector (TNT 12000 DoseMate)
2
2-19
5. Attach the triaxial cable to the chamber.
6. Use the cable clips to attach the cable down the rod.
W Caution
To prevent damage to the test equipment, attach the cable near
the bottom of the test stand. Make sure that the cable is not in
the X-ray beam between the tube and the ion chamber.
7. Put the test stand in the middle of the X-ray beam.
8. Attach the loose end of the triaxial cable to the input connector of the Dosimeter
Detector.
Note
To do an HVL test, put the HVL filters in the test stand filter tray.
Advertising