Dental waveform analysis, Dental with filament preheat (single phase), Dental without filament preheat (single phase) – Fluke Biomedical TNT 12000 User Manual

Page 123: Three phase, Hf (high frequency)

Advertising
background image

TNT 12000WD kVp, Exposure, and Exposure Time Measurement

Dental Waveform Analysis

E

E-3

Dental Waveform Analysis

In the Dental mode, waveform analysis of exposure time and kV measurements is done

as follows:

Dental with Filament Preheat (Single Phase)

Exposure time in milliseconds is calculated between the first and the last zero crossings

of the kV waveform plus 4 ms (compensates for the waveform time that does not produce

x-rays) after the preheat. Exposure time in pulses is measured between the first and the

last zero crossings of the kV waveform after the preheat.
Average kVp is measured on the pulses in the central 80 % of the kV waveform after the

preheat. Maximum kV and PPV are measured for the complete waveform after the

preheat. Exposure is measured from the start of the exposure to the end of the exposure.

Dental without Filament Preheat (Single Phase)

Exposure time in milliseconds is calculated between the first and last zero crossings of

the kV waveform plus 4 ms (compensates for the waveform time that does not produce x-

rays). Exposure time in pulses is measured between the first and last zero crossings of the

kV waveform.
Average kVp is measured on the pulses in the last 75 % of the kV waveform, less the last

peak. Maximum kV and PPV are measured for the complete waveform. Exposure is

measured from the start of the exposure to the end of the exposure.

Three Phase

Exposure time in milliseconds is calculated between the 75 % points on the rising and

falling edges of the kV waveform.
Average kVp is measured on the peaks in the central 80 % of the kV waveform.

Maximum kV and PPV are measured for the complete waveform. Exposure is measured

from the start of the exposure to the end of the exposure.

HF (High Frequency)

Exposure time in milliseconds is calculated between the 90 % points on the rising and the

falling edges of the kV waveform.
Average kVp is measured on peaks in the central 80 % of the kV waveform. Maximum

kV and PPV are measured for the complete waveform. Exposure is measured from the

start of the exposure to the end of the exposure.