Dosimeter measurements, How to add an ion chamber definition or cal factor – Fluke Biomedical TNT 12000 User Manual

Page 54

Advertising
background image

TNT 12000

Users Manual

2-26

Dosimeter Measurements

When the Display and Dosimeter Detector are connected and the ion chamber parameters

are set, the MEASUREMENT screen shows in the Display.

fct238.png

Figure 2-15. Dosimeter Measurement Screen

Note

Average mA, mAs, and Dose/mA are only shown when the mAs Detector is
connected to the Display.

With the ion chamber set in the X-ray beam, make an exposure. The Display and

Dosimeter Detector beep when an exposure is sensed and the Display is updated with the

measured values.
The procedure for all subsequent measurements depends whether the Automatic Reset

feature is turned on or not. See the Measurement Mode Selection section to set this

feature. If Automatic Reset is on, make an exposure and the Detector resets itself. If

Automatic Reset is off:
1. Use B or A to highlight the Reset button and push E.
2. When the reset message is removed from the display, make an exposure.
The MEASUREMENT screen is refreshed with new values after the exposure.

Note

When the mode is set to Dose Rate, the display refreshes at one second
intervals. See the Measurement Mode Selection section to set the mode.

How to Add an Ion Chamber Definition or Cal Factor

You can add a maximum of nine ion chamber definitions to the Detector from the

Dosimeter MEASUREMENT screen, or the ION CHAMBER SETUP screen. Each ion

chamber can have a maximum of 10 cal factors.
To add an ion chamber definition:
1. Use B or A to highlight the IC button in the Dosimeter MEASUREMENT screen and

push E. The ION CHAMBER SETUP screen shows in the Display.