Calibration and Verification

B

B-9

All 96035B ion chambers must have an actual correction factor within

±

1.5 % of the

value shown in the curves at any point.
The actual points plotted in the graphs below correspond to the points in the MV and MH

series of beam qualities. Users making measurements at a beam quality similar to one of

the PTB mammographic beams may easily obtain the appropriate correction factor by

assuming the HVL of the most equivalent PTB beam. For a more precise correction

factor value, the user may calculate the actual first HVL and locate the proper correction

value on the curve.

Note

Introduction of material, other than free air, behind the Ion Chamber will
cause its response to change due to backscatter.