Fluke Biomedical TNT 12000 User Manual

Page 116

Advertising
background image

TNT 12000

Users Manual

C-4

Table C-1. Specifications for IEC 61267 Defined Beam Qualities

kVp

Beam

Quality

First HVL

(mm Al)

Beam

Quality

First HVL

(mm Al)

Beam

Quality

First HVL

(mm Al)

40

RQR 2

1.42

RQA 2

2.2

50

RQR 3

1.78

RQA 3

3.8

60

RQR 4

2.19

RQA 4

5.4

70

RQR 5

2.58

RQA 5

6.8

80

RQR 6

3.01

RQA 6

8.2

90

RQR 7

3.48

RQA 7

9.2

100 RQR

8

3.97

RQA

8 10.1 RQT

8 6.9

120 RQR

9

5.00

RQA

9 11.6 RQT

9 8.4

150 RQR

10

6.57

RQA

10 13.3 RQT

10 10.1

500 – 100 Measurement ranges

Values for ion chambers are calculated using nominal sensitivities: 1 x 10

9

R/C.

Table C-2. 500-100 Measurement Ranges

Units Effective

Range

[1]

Threshold

Resolution

Step Size

R 10

μ

- 9999

250

μ

0.1

μ

R/s 10

μ

- 100

250

μ

0.1

μ

R/min

1 m – 6 k

15 m

0.01 m

R/hr

0.1 – 360 k

0.9

0.001

R/pulse

[2]

1 m – 1.67

4.17

μ

0.01

μ

Gy 0.1

μ

- 876 m

2.19

μ

0.001

μ

Gy/s 0.1

μ

- 876 m

2.19

μ

0.001

μ

Gy/min

0.01 m - 52.5

131

μ

0.1

μ

Gy/hr

1 – 3.15 k

7.88 m

0.01m

Gy/pulse

[2]

0.01

μ

- 14.6 m

36.5 n

0.1n

[1]

IEC 61674 effective range at 1 % resolution steps.

[2]

At 60 frames/second (1 to 120 frames/selectable).

Note

Introduction of material, other than free air, behind the Ion Chamber will
cause its response to change due to backscatter.