Operation

How to Use the Dosimeter Detector (TNT 12000 DoseMate)

2

2-23

4. Put the ion chamber bracket on the horizontal rod. Move it along the rod until the top

of the bracket aligns with the scribe mark in the middle of the rod. Tighten the

chamber bracket thumbscrew.

5. Put the HVL filter tray on the horizontal rod. Move it along the rod until the top of

the tray aligns with the top scribe mark on the rod. Tighten the HVL filter

thumbscrew.

6. Attach the ion chamber to the bracket.
7. Attach the triaxial cable to the chamber.
8. Use the cable clips to attach the cable down the rod..

W Caution

To prevent damage to the test equipment, attach the cable near

the bottom of the test stand. Make sure that the cable is not in

the X-ray beam between the tube and the ion chamber.

How to Place the Ion Chamber in Limited Access Situations

Use the plastic cable stem to accurately put the ion chamber in image-intensifier housings

or other hard to access situations. The plastic cable stem attaches to the triaxial-cable

connector nearest the ion chamber. This adds

five inches

of position control.

Triax Cable

Triax Cable Connector

96020C Ion Chamber

Cable Stem

Rod

fct27.eps

Figure 2-12. Ion Chamber Cable Stem

If more distance is necessary, attach the cable stem to the threaded end of the vertical rod,

as Figure 2-12 shows. This adds

15 inches

of rigid control.