Fluke Biomedical TNT 12000 User Manual

Page 45

Advertising
background image

Operation

How to Use the Dosimeter Detector (TNT 12000 DoseMate)

2

2-17

TNT 12000

DoseMate

OR

ION Chamber

Triax ION

Chamber Cable

ION Chamber

Input

mAs Input

TNT 12000

DoseMate

W/mAs

ION Chamber Input

fct25.eps

Figure 2-8. Ion Chamber Connection to Dosimeter

3. Connect the other end of the cable to the ion chamber input connector on the

Dosimeter Detector.

Note

For ion chamber protection, put the dust cover on the ion chamber
connector when the ion chamber is not connected to the Detector.

Put the ion chamber in the X-ray beam. Be sure the Detector and Display are not in the

radiation beam before the Dosimeter Detector and Display are turned on. A test stand and

adapter stem are included with the Dosimeter Detector. The sections that follow

demonstrate how the test stand helps hold the ion chamber in the X-ray beam.

How to Place the Ion Chamber for Over-Table X-ray Sources

When the X-ray source is above the table, use the test stand to hold the ion chamber in

the X-ray beam.
To install the ion chamber on the test stand, see Figure 2-9 and do the steps that follow: