Operation

How to Use the X-ray Detector (TNT 12000WD)

2

2-13

the highlighted value.

5. Use C or D to highlight the Continue button and push E to set the

parameters into the Detector.

Note

The parameter values set at this point redefines the profile selection.

The MEASUREMENT screen shows in the Display. The name on the middle button in

the bottom of the screen shows the profile (Auto, Manual, Default, or user-defined label)

to which the Detector is set.
To cancel parameter changes before the changes are set, highlight the Cancel button and

push E.
The Save As button only shows in the MANUAL settings screen and makes a user-

defined profile. When the profile is a user-defined profile, the Save As button is

replaced with a Save button. The Save button allows you to change the name of a user-

defined profile. See the How to Make a User-Defined Profile section for more on user-

defined profiles.

How to Make a User-Defined Profile

User-defined profiles, or parameter sets, are specified through the Manual setting screen.
To make a user-defined profile:
1. Select a measurement mode in the MODE SETUP screen.
2. Use C or D to highlight the Settings button and push E to open the

MEASUREMENT SETTINGS screen.

3. Use C or D to highlight the Manual button and push E.
4. Use C or D to highlight the Edit button and push E.
5. Use C or D to highlight the parameter you want to change and push E to open

the drop-down list.

6. Use C or D to highlight a value in the list and push E to set the parameter to

the highlighted value.

Repeat steps 5 and 6 for each parameter that needs to be changed.

7. Use C or D to highlight the Save As button and push E to show the

ALPHANUMERIC SCREEN in the Display.

fct33.bmp