Detector placement with below table x-ray source, How to measure x-ray parameters – Fluke Biomedical TNT 12000 User Manual

Page 35

Advertising
background image

Operation

How to Use the X-ray Detector (TNT 12000WD)

2

2-7

Note

Set the Detector so all of the active area of the Detector is exposed to the X-
ray beam. A

22-inch

distance from the focal point to the Detector is

suggested.

Detector Placement with Below Table X-ray Source

For X-ray machines that have the X-ray source below the table, align the Detector face

down, in the middle of the beam, with the long axis of the Detector perpendicular to the

axis of the X-ray tube. See Figure 2-5.

Under Table

X-Ray Tube

Tube Axis

TNT 12000WD

Long Axis of TNT 12000WD

X-Ray Table

fct06.eps

Figure 2-5. X-ray Detector Setup for Below Table X-ray Source

How to Measure X-ray Parameters

Connect and set up the X-ray Detector for a measurement. See the How to Setup the

System section above for connection and set up instructions. The MEASUREMENT

screen in Figure 2-6 shows in the Display.

Note

If the mAs Detector is already connected, an mAs Measurement Option
screen can show in the display. Go to the How to Use the mAs Detector
manual section to set the mAs measurement options. When done, come back
to this section to continue the X-ray measurement.