How to make a dosimeter measurement, Ion chamber setup – Fluke Biomedical TNT 12000 User Manual

Page 52

Advertising
background image

TNT 12000

Users Manual

2-24

How to Make a Dosimeter Measurement

After the ion chamber is connected to the Detector, push O on the Detector and Display.

Note

If the mAs Detector is already connected, an mAs Measurement Option
screen can show in the display. Go to the How to Use the mAs Detector
manual section to set the mAs measurement options. When done, come back
to this section to continue the Dosimeter measurement.

Once connected to the Dosimeter Detector, the system checks if an ion chamber is set for

the measurement. If set, the MEASUREMENT screen in Figure 2-15 shows in the

Display. If an ion chamber is not set, then the IC NOTIFICATION screen shown in

Figure 2-13 shows in the Display. Go to the Ion Chamber Setup section if an ion chamber

is necessary for the measurement.

fct240.png

Figure 2-13. Ion Chamber Notification Screen

Select OK and push E to select or configure an ion chamber. If you select Cancel,

Dose and Dose Rate are set to C and A, and you cannot change the measurement units.

Ion Chamber Setup

Open the ion chamber setup screen opened from the IC NOTIFICATION screen or the

Dosimeter MEASUREMENT screen.
To setup the TNT 12000 for an ion chamber:
1. Use B or A to highlight OK in the IC NOTIFICATION screen or the IC button in the

Dosimeter MEASUREMENT screen.

2. Push E to open the ION CHAMBER SETUP screen shown in Figure 2-14.