Ptb calibration option, Other calibration options, Calibration accuracy – Fluke Biomedical TNT 12000 User Manual

Page 106: Calibration verification, Energy correction factors

Advertising
background image

TNT 12000

Users Manual

B-4

Diagnostic Unattenuated beam ........................... Calibration is performed at the Fluke Biomedical, Radiation

Management Services equivalent of DV70 (PTB defined as 70 kVp,
first HVL of 2.45 mm Al).

Mammographic beam ........................................... Calibration is performed at the Fluke Biomedical, Radiation

Management Services equivalent of MV30 (PTB defined as 30 kVp,
first HVL of 0.337 mm Al).

PTB Calibration Option

This calibration option replaces the standard calibration when PTB requirements must be

met, or when specifically requested by a customer. This option includes two diagnostic

points and one mammographic point as given below. Calibration factors are normalized

to 20

°

C and 1013 hPa of atmospheric pressure.

Diagnostic Unattenuated beam ........................... Calibration is performed at the Fluke Biomedical, Radiation

Management Services equivalent of DV70 (PTB defined as 70 kVp,
first HVL of 2.45 mm Al).

Diagnostic Attenuated beam ............................... Calibration is performed at the Fluke Biomedical, Radiation

Management Services equivalent of DH70 (PTB defined as 70 kVp,
first HVL of 6.77 mm Al).

Mammographic Beam........................................... Calibration is performed at the Fluke Biomedical, Radiation

Management Services equivalent of MH30 (PTB defined as 30 kVp,
first HVL of 0.67 mm Al).

Other Calibration Options

Optional calibration points are the Fluke Biomedical, Radiation Management Services

equivalent of M50, S75, H60, MoMo35, DV30, DV40, DV50, DV60, DV80, DV90,

DV100, DV120, DV150, DH40, DH50, DH60, DH80, DH90, DH100, DH120, and

DH150. Calibration factors at these points may be normalized to either 20

°

C or 22

°

C, as

specified by the customer.

Calibration Accuracy

NIST Traceable ...................................................... NIST traceable reference chambers are accurate to within

±

1 % at

each beam quality. Matching between the reference and unknown
chamber is accurate to

±

1 %. Thus, NIST traceable calibrations are

accurate to within

±

2 %.

PTB traceable ........................................................ PTB traceable reference chambers are accurate to within

±

1.5 % at

each beam quality. Matching between the reference and unknown
chamber is accurate to

±

1 %. Thus, PTB traceable calibrations are

accurate to within

±

2.5 %.

Calibration Verification

NIST traceable ....................................................... For NIST traceable calibrations, a chamber will be reported to be

outside of its calibration accuracy specification when a new calibration
value differs from the old calibration value by more than

±

4 %. This

includes the

±

2 % accuracy specification plus

±

1% for the verification

reference plus

±

1 % for the verification match.

PTB traceable ........................................................ For PTB traceable calibrations, a chamber will be reported to be

outside of its calibration accuracy specification when a new calibration
value differs from the old calibration value by more than

±

5 %. This

includes the

±

2.5 % accuracy specification plus

±

1.5 % for the

verification reference plus

±

1 % for the verification match.

Energy Correction Factors

Diagnostic – Unattenuated Beam (In Front of the Phantom)

The energy correction factors for the 96035B are determined for the unattenuated

diagnostic beam using the PTB defined DV series of beam qualities given below in Table

B-1.

Advertising