Test 10.1.9 - SCSI CDB – LOG SENSE

Purpose: To determine that the DUT, an SSP Target port, properly handles a LOG SENSE command.

References:

SPC-3 6.6

Resource Requirements: SAS Protocol Analyzer and Generator.

Revision History: September 28, 2004

Test Setup: The DUT and the Testing Station are physically connected.

Test Procedure (SSP Target):

• The Testing Station should start and complete a Phy Reset sequence followed by an Identify sequence

with the DUT. Since the DUT is a target the Testing Station should transmit an Identify Address frame
indicating that it is a SAS Initiator.

• The Testing Station should open an SSP connection to the DUT and transmit a SCSI LOG SENSE

command with the PPC and PARAMETER POINTER fields set to 0, a LOG PAGE CODE of 00h
(Supported Log Pages), and SP bit of 0. Close the connection

• The DUT should open a connection to the Testing Station and transmit DATA frames to the DUT.

• Allow the DUT to open an SSP connection and transmit a SCSI RESPONSE frame to the received

command.

Observable Results: Verify that the DUT transmits RESPONSE frame indicating that the command completed
with GOOD status. Verify that the DUT transmitted Log Page data formatted according to 7.2.12 of SPC-3. Verify
that the DUT reported its supported Log Pages defined in table 194 of SPC-3.