Teledyne LeCroy SAS Verification Test Descriptions User Manual

Page 81

Advertising
background image

The University of New Hampshire

InterOperability Laboratory



Serial Attached SCSI Consortium

80

SAS Link and Transport Layer Test Suite v0.13

Test 7.4.3 - Closing SSP Connections – DONE (CREDIT TIMEOUT)

Purpose: To determine that the DUT properly responds when RRDY has not been received for an impending
transaction.

References:

7.16.6 SAS Standard

Resource Requirements: SAS Protocol Analyzer and Generator

Revision History: September 13, 2004

Test Setup: The DUT and the Testing Station are physically connected.


Test Procedure (Initiators):

• The Testing Station should start and complete a Phy Reset sequence followed by an Identify sequence

with the DUT. Since the DUT is an initiator the Testing Station should transmit an Identify Address
frame indicating that it is a target.

• Wait for the DUT to open an SSP connection to the Testing Station. The Testing Station should not

transmit RRDY to grant credit for the DUT to transmit a SCSI Command to the Testing Station.


Observable Results: Verify that the DUT transmitted DONE (CREDIT TIMEOUT) within 1 msec of opening the
SSP connection.

Advertising